Material-Surface-Metall & Sheet Testers

MR5-200C MICRO-OHMMETER

featuring continuous measurement current up to 200 A

The MICRO-OHMMETER MR5-200C is used to determine electric resistances on very low – resistance objects (e. g. bus bars, any connectors) while using a very high, continous flowing measurement current. The MR5-200C is able to generate a stable maximum current up to 200 A.

Under normal conditions, the electric resistance can be determined using a much lower current. For example, our MICRO-OHMMETER MR300C-A has a 10A measurement current with a resolution of 10nW (10-8W). However, faulty connections can not be found this way, as a high current is needed to determine the dissipation causing unstable or rising measurement values on bad connections.

To perform this a high measurement current of the MR5- 200C can be generated constantly over longer periods. Additionally, the measurement values can be send to any PC based spreadsheet program e.g. EXCEL© using the integrated interface. Time based sending is also supported, e.g. when evaluating the dissipation, measurement values are sent via interface with one second time gap.

TEST CURRENTmax. 200A (True DC). Increment 1 A.
Adjustable and presetable from 0 to 100%
0-199.9 µΩ (test current: up to 200 A).

0-1.999 mΩ (test current: up to 200 A).

0-19.99 mΩ (test current: up to 100 A).

0-199.9 mΩ (test current: up to 20 A).

0-1999 mΩ (test current: up to 2 A).
READING RESOLUTION0.1 μΩ for R < 200 μΩ.

1 μΩ for R < 2000 μΩ.

10 μΩ for R < 20 mΩ.
OUTPUT VOLTAGE4..5 Vdc @ all currents.
DISPLAY4 digits result, current and time
LCD 9mm character height with backlight
TOTAL RESISTANCEmax. 20 mΩ @ 200A.

max. 40 mΩ @ 100A.
MEASUREMENT PRINCIPLEFour-terminal due to Kelvin
Protects the instrument from overheating.
CONTINUOUS OPERATION TIMEAt 200 A this device should not be used continously for more than 30 minutes (overheat protected). At 100 A the measuring time is unlimited.
BASIS ACCURACY± 0.3 % of reading ± 0.1 % of Range.
CURRENT FLOW DURATIONThe duration (time) for steady current flow can be easily set to 1,2,5,10,20s. ....up to 600s or unlimited (o.h. protected).
ADVANCED FEATURESDigital direct reading of resistance, current, measuring time and updated every second
BUILT-IN MEMORYStorage of 1000 test readings.
INTERFACERS232 @ 2400, 4800, 9600 baud. , USB-B
ENVIRONMENTAL
PROTECTIONS
IP54 (with closed lid).
SAFETY CLASSMeets the requirements of IEC 61010-1:1990, IEC 61010-1:1992 amendment 2.
POWER SUPPLY90~264 VAC - 47~63Hz / 127~370VDC
OPERATING TEMPERATURE-5°C to 50°C.
STORAGE TEMPERATURE-40°C to 85°C.
HUMIDITY RANGE10~95 % of RH (non condensing).
EQUIPMENT WEIGHTApprox. 9.5 kg. (without cables)
DIMENSIONS410W x 330D x 180H mm.
PRINTERThermal printer

All technical datas are evaluated with 5 m Standard Measuring Cables, 100 µΩ test resistance, 230V line source, at 23°C room temperature.

This instrument can be delivered as MR5-200P version too. There are two main differences between the MR5-200C and the MR5-200P:

The C-Version uses continuous measuring current, the P-Version offers only a current duration of one second maximal. On the other hand the P-Version can be used for short time measurements without line connection, the C-Version needs line source for all operation modes.

MR5-200CMR5-200P
Current flowup to 1 h cont.3 s at 200 A
Battery operationnoyes
Max. Measurementsno limit100 @ 200A battery only
Weight9.5 kg10.5 kg
Inductive objectsonly small loadsno inductive loads


All other technical datas and features are identical.

DIN EN ISO 9001:2008

Fragen?

Tel.: +49 (0)3328 / 3179 – 0
Fax: +49 (0)3328 / 3179 – 10
E-Mail: sales@schuetz-messtechnik.com

Hier erhalten Sie Hilfe bei technischen Fragen und weitere Informationen über Preise, Versand und Vertrieb

MR-1